How to use nucfilms actinides adsorbing sheets
PRELIMINARY
Warning
: The method described is recommended only for untreated waters having
drinking water quality. For waste waters the method may be used, but requires
special precaution.
Nucfilm s actinides adsorbing sheets are polymethylmetacrylate sheets
covered on one side with a binder holding a finely ground cation exchanger
(Diphonix ,Eichrom Industries).
Sampling
There is no need to acidify waters having drinking water quality. Simply avoid
glass containers.
Use polyethylene bottles and fill them completely.
Exposition to the water sample
Fill 80 ml of your untreated water sample into a 100 ml FEP beaker. Don't use
glass beakers.
FEP (Fluor-Ethylene-Propylene) showed negligible adsorption. Add 0.5 ml
concentrated
formic acid (85%) to reduce
pH to 2 to 3
. If your sample is already at low pH, e.g. a
nitric acid solution, adjust pH to around 2 and just try if it works.
Rinse the sheet with distilled or deionized water and fix the sheet with a
stainless steel
clamp or glue it to a plastic (e.g. Plexiglass) holder (Pattex
works well as a glue).
Place it in your sample as shown in the
figure
. Stir at approx. 200 rpm using a teflon
covered stirring rod. After
20 to 24 h
remove the sheet from your sample, rinse it with
distilled or deionized water and dry it with not too hot air (hair drier).
Dont touch the surface.
For some samples containing high CO
2
concentrations
gas bubbles may form on the film surface. In this case just lift the sheet
shortly out of the sample and put it back.
This removes the bubbles. This prodcedure may be necessary several times during
the first hour of exposition.
It is not known if these gas bubbles hinder adsorption.
Any data would be welcome
.
Measurement
With the sheet at a
distance of about 10 mm
from the surface of a 900 mm
2
alpha detector you have a detection efficiency around 20% .
Assuming a 100% adsorption efficiency
1 count/1000s thus corresponds to around 65 mBq/l in your sample.
Adsorption efficiency and calibration
For drinking waters you can expect to have adsorption efficiencies > 80% for
20 mm x 20 mm
sheets exposed as given above. To verify adsorption efficiency you can expose
successively
several sheets in the same 80 ml sample. Assuming that every sheet takes up the
same fraction
of the uranium that remained in the sample, adsorption efficiency will be 1- (U
activity on sheet n+1)/
(U activity on sheet n).
U standard solutions can be prepared with uranylnitrate.
Energy resolution
Film thickness is several µm making energy resolution far worse than for
the
Ra adorbing films. However even when measured in air you still have
234
U and
238
U
separated. There is an interference of the
234
U lines with the
226
Ra lines.
Actually Ra adsorbtion efficiency of these films seems to be far lower than
U adsorbtion
efficiency. So far there has been no systematic tests on Ra adsorbtion. Thus
any data would be welcome
.
December 24, 2000 Heinz Surbeck